Development and Testing of a Data Capture Device for Use With Clinical Incentive Spirometers: Testing and Usability Study

Development and Testing of a Data Capture Device for Use With Clinical Incentive Spirometers: Testing and Usability Study

Development and Testing of a Data Capture Device for Use With Clinical Incentive Spirometers: Testing and Usability Study

Original Paper

1Department of Anesthesiology, University of Michigan Health System, Ann Arbor, MI, United States

2Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI, United States

Corresponding Author:

Michael L Burns, MD, PhD

Department of Anesthesiology

University of Michigan Health System

1500 E Medical Center Drive

Ann Arbor, MI, 48109

United States

Phone: 1 734 936 4171

Email: mlburns@med.umich.edu


Background: The incentive spirometer is a basic and common medical device from which electronic health care data cannot be directly collected. As a result, despite numerous studies investigating clinical use, there remains little consensus on optimal device use and sparse evidence supporting its intended benefits such as prevention of postoperative respiratory complications.

Objective: The aim of the study is to develop and test an add-on hardware device for data capture of the incentive spirometer.

Methods: An add-on device was designed, built, and tested using reflective optical sensors to identify the real-time location of the volume piston and flow bobbin of a common incentive spirometer. Investigators manually tested sensor level accuracies and triggering range calibrations using a digital flowmeter. A valid breath classification algorithm was created and tested to determine valid from invalid breath attempts. To assess real-time use, a video game was developed using the incentive spirometer and add-on device as a controller using the Apple iPad.

Results: In user testing, sensor locations were captured at an accuracy of 99% (SD 1.4%) for volume and 100% accuracy for flow. Median and average volumes were within 7.5% (SD 6%) of target volume sensor levels, and maximum sensor triggering values seldom exceeded intended sensor levels, showing a good correlation to placement on 2 similar but distinct incentive spirometer designs. The breath classification algorithm displayed a 100% sensitivity and a 99% specificity on user testing, and the device operated as a video game controller in real time without noticeable interference or delay.

Conclusions: An effective and reusable add-on device for the incentive spirometer was created to allow the collection of previously inaccessible incentive spirometer data and demonstrate Internet-of-Things use on a common hospital device. This design showed high sensor accuracies and the ability to use data in real-time applications, showing promise in the ability to capture currently inaccessible clinical data. Further use of this device could facilitate improved research into the incentive spirometer to improve adoption, incentivize adherence, and investigate the clinical effectiveness to help guide clinical care.

JMIR Biomed Eng 2023;8:e46653

doi:10.2196/46653

Keywords



Pulmonary complications after major surgery, including pneumonia, atelectasis, respiratory failure, prolonged supplemental oxygen requirements, and reintubation, are common, expensive, and harmful to patients [1]. Studies estimated these complications in the range of 2%-39% [2,3], though atelectasis alone has been found to affect up to 92% of postsurgical patients [4]. Originating in the 1970s, the incentive spirometer was designed to mimic the physiology of a sigh or a yawn—a slow voluminous inhalation [5,6], and this basic medical device is often used in postoperative care to aid lung expansion to prevent or reduce respiratory complications.

To complicate the matter, correct use and device adherence is low among patients. Guidelines on how to properly use the incentive spirometer device outline best practices [7,8], but over 26% of patients fail to use their incentive spirometer correctly, and over 38% deny ever using their device in their postoperative care [9], highlighting the need for more evidence-based recommendations. There is sparse evidence for the use of the incentive spirometer device [10] for postoperative pulmonary complication prevention [11] with only a few studies demonstrating clinical effectiveness when used properly [12,13]. As a result of the lack of high-level evidence, some clinical practice guidelines do not support its routine postoperative use [14]. Additionally, there remains disagreement as to the most effective way to use the device, as studies have been unable to demonstrate its superiority over other techniques such as deep breathing techniques, directed coughing, early mobilization, and optimal analgesia [15]. Uncertainty around the effective spirometry use is partially due to the scarcity of spirometer compliance data [16]. Compliance measurements, made through self-reporting and staff observation, are difficult to obtain, and when captured, they have demonstrated low patient adherence to the incentive spirometer device [12]. Though data remain elusive, 86% of health care providers believe patient adherence is poor, and 95.4% believe it should be improved [17], demonstrating the perceived use of incentive spirometry. The first step in determining the optimal use of the incentive spirometer is to improve data collection. Automated capture of spirometry data may improve the quality of research studies and ultimately determine the incentive spirometer’s use in improving lung function and minimizing postoperative complications.

Incentive spirometers are simple plastic meters to measure inhalational breath flow and volume; they lack the ability to record data. Digital flowmeters can be used to replicate the incentive spirometer [18]. While an option for improving data capture, the digital flowmeter can be clinically infeasible in most practices due to its complexity and cost. Collecting data directly from incentive spirometers falls into a technology category called the “Internet-of-Things” [19]. There have been great advances in the miniaturization of computing devices and the evolution of the “Internet-of-Things” into mainstream health devices [20,21]. As a result, wearable technologies and web-based platforms are capturing more clinical data now than ever before [22,23]. These technologies give clinicians and researchers access to otherwise inaccessible patient data and the ability to investigate new data interactions, such as enabling patients to exercise with wearable device hardware (eg, chest monitors and watches) and incorporating and communicating these data with patient physiologic and movement data [24,25]. The goal was to use this technology for the incentive spirometer.

We hypothesize an add-on device can accurately measure flow and volume data from a common incentive spirometer. This paper describes the creation and testing of an incentive spirometer add-on device to measure flow and volume data. An Internet-of-Things approach was adopted to enable this device to work with an existing incentive spirometer to capture physiological data and communicate externally on a closed private internet connection using hospital Wi-Fi and Bluetooth technologies. Captured data were tested for use by developing a classification algorithm to determine valid from invalid breaths.


Ethics Approval

Institutional review board approval was obtained for this prospective exploratory study on April 15, 2021 (HUM00196543, University of Michigan, Ann Arbor). Informed consent was waived as participants were limited to authors within this study, and interventions for this study are limited to existing approved uses of the incentive spirometer.

Add-On Device Creation

An add-on device was designed and created for use with the Hudson RCI Voldyne 5000 Volumetric Exerciser (Teleflex Medical) incentive spirometer (Figures 1-3 and Multimedia Appendix 1) and composed of photoelectric reflective infrared optical sensors (Xingyheng) positioned lateral to the spirometer volume and flow columns. Ten sensors were placed along the spirometer volume column with each sensor along a 500 mL marking on the incentive spirometer device (500-5000 mL). Three sensors were placed along the flow column corresponding to the middle of flow spirometer markings (“best,” “better,” and “good”). Sensors were connected to a microcontroller by a solderless breadboard, using an ESP32 development board (Dorhea), a low-cost low-power system-on-a-chip microcontroller with integrated Wi-Fi and dual-mode Bluetooth wireless communication capability. Components were soldered on a breadboard with breadboard jumper wires and resistors connecting components. Interactions between the microcontroller and the sensors were coded using C++ (Bell Laboratories of American Telephone and Telegraph). Components were encased in a 3D-printed base situated beneath the incentive spirometer with a layer of plexiglass fixed overtop.

Data were sent directly to an iOS application running on an iPad Pro (12.9″ and 10.2″, fifth generation, Apple) through ESP32 Bluetooth as well as stored on the add-on device in a 32 gigabyte microsecure digital card (Kootion) and card reader module (HiLetGo). A clock module (Melife) time-stamped data. Further interconnection was made possible using an analog digital multiplexer breakout board (Xie QianJin) and included a 3.7-V 2400-mAh rechargeable lithium battery (Akzytue) and charging module (MakerFocus) to power the device. An alternative design for the add-on base can be found in Multimedia Appendix 2.

Figure 1. The incentive spirometer piston positions were read using reflective infrared optical sensors (1) and relayed to an ESP32 in the base of the add-on device (2). The device both stored the data internally and used Bluetooth technology to transmit spirometer data to an Apple iPad (3). Data were then transmitted from the iPad to dedicated servers for further data processing and storage (4).
Figure 2. Ten photoelectric reflective infrared optical sensors were placed along the spirometer volume column with each sensor along a 500 mL marking on the incentive spirometer device (500-5000 mL); 3 sensors placed along the flow column. An on/off switch is featured on the front of the device. An ESP32 development board, rechargeable lithium battery, and real-time clock module are labeled. Connections between components were made with breadboard jumper wires and resistors, and all components are encased in a 3D-printed base with a layer of plexiglass overtop. A microsecure digital card reader module is in the back of the base (not shown).
Figure 3. The Hudson RCI Voldyne 5000 Volumetric Exerciser (Teleflex Medical) incentive spirometer is situated over the base and locked into place with an overlying base lip and a screw on top of the 3D-printed posterior arm. A charging port is located in the base (right picture).

Software and Video Game Creation

An iOS video game and analytics application was built using Unity and C# (Microsoft). A web server receives the data and handles application programming interface (API) requests written in Java (Sun Microsystems). The ESP32 reads the sensor data using custom C++ code and sends processed data to an Apache web server via Wi-Fi through an iPad device connected via Bluetooth (Figure 1). Python (Python Software Foundation) applications were exposed by the web server, and downstream applications were networked with the server through an API to allow data collection and further use. A web server receives the data and handles API requests. A video game was developed specifically for use with the incentive spirometer and add-on device serving as a controller, designed for use with an Apple iPad Pro.

Add-On Device Sensor Testing

Two investigators (AS and MLB) tested the add-on device without crossover using the Hudson RCI Voldyne 5000 Volumetric Exerciser incentive spirometer. In these tests, 5 user breaths were attempted at each of the 10 volume sensor positions (100-5000 mL) for a total of 50 breaths per user. The volume goal was to get the top of the volume piston to the desired volume marking, while the flow goal was to get the top of the flow bobbin to the middle of the desired flow marking. Flow readings were also tested with every volume test with an additional 5 flow tests at each of the 3 flow sensor positions. To attain the desired flow or volume piston levels, users were allowed to breathe through or tilt the incentive spirometer device.

A single investigator (MLB) tested flow and volume sensor ranges with the incentive spirometer connected to the Puritan-Bennet PTS-2000 Ventilator Analyzer Tester (Mallinckrodt) digital flowmeter. In this testing, the investigator attempted increasing volume and flow values to identify sensor ranges. Volume measurements from the flowmeter were corrected using the body temperature, pressure, water vapor saturated method (correcting for body temperature [37 °C], ambient pressure, and gas saturated with water vapor). The flowmeter volume was calculated to be 183.75 mL by calibrating the flowmeter readings precisely at the 1000 mL display level of the Voldyne 5000, averaged across 10 breaths. This volume was subtracted from each raw volume reading from the flowmeter. Flowmeter testing was attempted with a minimum of 25 breaths around each flow and volume sensor level, independently, and the volume achieved from the flowmeter and the sensor level attained from the iPad were recorded. Due to limitations continually achieving breath volumes above 3250 mL (sensor level 7), additional testing using the 2500 mL volume Hudson RCI Voldyne 2500 Volumetric Exerciser (Teleflex Medical) incentive spirometer was conducted. The add-on device was created with each sensor aligning to the 500 mL markings of the Voldyne 5000. When using the Voldyne 2500, the sensors were near but incompletely aligned to the 250-mL markings.

Breath Algorithm Development and Testing

While sensors in the add-on device can detect where a piston is located, they cannot determine if a breath was conducted to achieve the sensor level. It is important to be able to accurately identify when a breath moves the incentive spirometer piston and bobbin as opposed to tilting the incentive spirometer causing them to fall to the desired level. To aid in distinguishing a breath attempt, an algorithm was developed in Python (version 3.1) to process and classify each user's breath using flow and volume data over time. This algorithm reads spirometer log data into a pandas data frame and parses data by identifying the start and end of breaths. Breath start was identified by a zero flow value that precedes a positive flow value, while breath end is identified by 2 consecutive zero flow values. Once identified, each breath is classified as “valid” or “invalid.” A valid breath must have the following volume criteria: volume starts at 0, increases within 1.5 seconds of breath start, >0 throughout the breath, does not decrease while there is positive flow, and the length of the breath is between 2 and 15 seconds. Breaths that do not meet these requirements were deemed invalid.

To test this algorithm, separate from device testing, 2 investigators (AS and MLB) each attempted to create 5 valid and 5 invalid breaths at each volume sensor level. These breath data were processed through the classification algorithm and evaluated. Invalid breaths were created by manually tipping the incentive spirometers or starting breaths at a starting position >0. All data validation was retrospectively validated using expert opinion from one of 2 users (AS and MLB).

Statistical Analysis

Summary and descriptive statistics were calculated for collected data using basic statistical techniques to assess models created for breath analysis including accuracy (as defined by user identification as gold standard), as well as mean, median, and SD for volume and flow measurements in sensor range testing.


Add-On Device Sensor Testing

An add-on incentive spirometer device was created to independently measure real-time spirometer flow and volume piston positions (Figures 1-3). From 2 investigators, 99 of 100 volume readings were measured at the correct corresponding volume sensor (accuracy 99%, SD 1.4%). Only a single volume reading was inaccurate, failing once to capture at the lowest volume sensor (500 mL). All 130 flow bobbin readings corresponded to the correct flow sensor (100% accuracy).

Volume and flow sensor ranges were determined using a digital flowmeter in line with the 5000 and 2500 mL incentive spirometers independently (Figure 4 and Tables 1 and 2). Add-on sensors were designed with placement at every 500 mL marking on the Voldyne 5000 incentive spirometer. These sensor placements were not adjusted for the 2500 mL incentive spirometer testing. Investigator breath limitations were reached at volume levels approaching 4000 mL. Between the 2 devices tested, median and average volumes were within 7.5% (SD 6%) of target volume sensor levels. Maximum triggering values seldom exceeded the intended sensor level, showing a good correlation to placement. Flow levels corresponding to “best,” “better,” and “good” levels differed significantly between devices with the 5000 mL flow values more than double the 2500-mL incentive spirometer at each level.

Figure 4. User testing of add-on device connected to a digital flowmeter for both the Voldyne 5000 (A) and the Voldyne 2500 (B) incentive spirometers. Circles represent individual breath attempts with digital flowmeter volume readings (L, x-axis) and add-on device sensor iPad readings (y-axis). Sensor levels were designed to be placed at 500 mL levels of the Voldyne 5000 incentive spirometer and closely aligned to 250 mL markings of the Voldyne 2500.
Table 1. Volume sensor ranges.a
NumberBreaths, nVolume (L)


Mean (SD)Median (IQR)MinimumMaximum
Voldyne 5000 volume sensor

1430.61 (0.10)0.62 (0.54-0.67)0.430.92

2710.98 (0.20)0.99 (0.81-1.13)0.631.55

3741.55 (0.18)1.54 (1.40-1.68)1.271.92

4602.20 (0.12)2.17 (2.11-2.28)1.992.44

5882.59 (0.20)2.59 (2.42-2.76)2.212.98

6373.09 (0.13)3.10 (3.02-3.16)2.793.38

7483.59 (0.35)3.53 (3.29-3.96)3.004.18

814.29 (N/Ab)4.29 (N/A)4.294.29

90N/A (N/A)N/A (N/A)N/AN/A

100N/A (N/A)N/A (N/A)N/AN/A
Voldyne 2500 volume sensor

1130.21 (0.04)0.22 (0.18-0.23)0.130.28

2350.41 (0.13)0.43 (0.29-0.48)0.220.69

3210.73 (0.11)0.72 (0.64-0.82)0.570.92

4291.03 (0.09)1.02 (0.99-1.09)0.871.20

5231.26 (0.10)1.27 (1.18-1.34)1.091.43

6241.60 (0.11)1.61 (1.53-1.67)1.381.78

7161.89 (0.11)1.91 (1.81-1.96)1.712.05

8252.21 (0.11)2.24 (2.13-2.28)1.992.38

9262.47 (0.11)2.49 (2.40-2.55)2.242.63

10372.75 (0.14)2.74 (2.64-2.85)2.523.09

aResults from investigations of increasing volume sensor readings from the Voldyne 2500 and Voldyne 5000 incentive spirometers. The number of breaths (n) at each sensor level (1-10) and the average, median (IQR), minimum, and maximum readings were obtained from a digital flowmeter in liters (L).

bN/A: not applicable.

Table 2. Flow sensor ranges.a

Breaths, nFlow (L/minute)


Mean (SD)Median (IQR)MinimumMaximum
Voldyne 5000 flow sensor

Best3026.36 (5.56)26.04 (22.51-27.76)19.2140.21

Better3044.31 (5.49)43.56 (40.44-48.25)36.1056.69

Good3157.75 (7.74)58.24 (53.99-61.27)39.3671.38
Voldyne 2500 flow sensor

Best3012.24 (2.36)12.31 (10.69-13.12)8.4319.63

Better3018.70 (1.92)19.08 (17.45-19.56)15.0722.60

Good3128.45 (3.34)28.13 (25.93-29.96)22.6138.30

aResults from investigations of increasing flow sensor readings from the Voldyne 2500 and Voldyne 5000 incentive spirometers. The number of breaths (n) at each sensor level (“best,” “better,” and “good”) and the average, median (IQR), minimum, and maximum readings were obtained from a digital flowmeter in liters per minute (L/minute).

Breath Algorithm Development and Testing

After data collection, an algorithm was developed to classify breath data to determine when the spirometer was actively used. This algorithm works by classifying each individual breath using criteria to determine their validity as described in the Methods section. Independent of sensor testing, 2 members of the study team attempted 5 valid and 5 invalid breaths at each of ten 500-mL volume levels (500-5000 mL). Investigator limitations in achievable volumes resulted in a total of 65 valid breaths and 100 invalid breaths. The breath classification algorithm resulted in a 100% sensitivity and a 99% specificity for the classification of “valid” versus “invalid” user breaths. A single valid reading was inappropriately classified by the algorithm, occurring at the 500 mL volume sensor level. Example breath algorithm results are shown in Figure 5.

To investigate the downstream applicability of breath data, a video game was developed using Unity software, Apple iOS, with a single final application installed on Apple iPad devices (Figure 6). The game was a Kirby (Nintendo)-based side scroller, where the character would approach an obstacle and traverse the obstacle after a successful breath. This game was developed specifically for use with the incentive spirometry device with game-play centered around proper use [7,8]. The incentive spirometer with an add-on device was successfully used to control the created game. A video showing the game played in real time with an incentive spirometer device can be found in Multimedia Appendix 3.

Figure 5. Example results from the breath algorithm used to classify “valid” from “invalid” breath attempts. The y-axis represents sensor levels for flow and volume readings in the incentive spirometer device: flow (0-3; corresponding to none: 0; “good”: 1; “better”: 2; and “best”: 3); volume (0-10, representing each 500 mL increment from 0 to 5000 mL). The x-axis represents the time from breath start in seconds.
Figure 6. A video game was specifically designed for use with the incentive spirometer and add-on device based on the Kirby Nintendo character. A sprite sheet for the character development within the video game is shown on the left with a screenshot of the video game shown on the right. Multimedia Appendix 3 shows the game being controlled using an incentive spirometer.

Principle Findings

In this study, an add-on device was created to allow the capture of incentive spirometry data with high accuracy (99%, SD 1.4% volume and 100% flow when tested at incentive spirometer level markings). Maximum triggering values were rarely exceeding intended sensor levels, proving excellent differentiation of levels in the design. Furthermore, while sensor placements were designed around the Voldyne 5000 incentive spirometer, the similarly shaped Voldyne 2500 performed well in testing without sensor modifications, suggesting use of the add-on device to similar incentive spirometers without significant redesign. While range testing was not intended to study volume sensor accuracy, between the 2 spirometers tested, median and average volumes were within 7.5% (SD 6%) of target volume levels, with the worst individual readings at lower volumes (250-500 mL sensors) with readings maximally 40 to 120 mL off target, respectively.

To differentiate quality breath attempts from errors, a classification breath algorithm was developed. The device allowed identification of start and end of breath attempts, valid from invalid breaths assessment, maximum flow, maximum volume, and volume or time ascents and descent calculations. The breath classification algorithm used device data to discern valid versus invalid breath attempts, showing 100% sensitivity and 99% specificity. Identifying valid breath data from noise is critically important for downstream applications and, while the classification algorithm yielded results, further efforts for improvement could be made using additional rule-based systems or machine learning algorithms.

This system is intended to enable clinical providers access to previously inaccessible spirometry data to improve spirometer instruction and use protocols, study patient compliance, and incentivize use. Using an add-on device similar to what was created in this study would increase the granularity of spirometer compliance data and could be used to provide insight into proper incentive spirometer use. Additionally, the add-on device can allow focused interventions to improve adherence. Reminder notifications alone have been shown to improve incentive spirometer use. In one study, an add-on use-tracking device was equipped with a bell that sounded for up to 2 minutes every hour as a reminder for the patient to use their incentive spirometer [12]. This study demonstrated that patients using the reminder device had a greater number of mean daily inspiratory breaths and a percentage of recorded hours with an inspiratory breath event. More importantly, patients with the reminder displayed significantly lower mean atelectasis severity scores measured by chest radiography, reduced median postoperative and intensive care unit length of stay, and had a lower mortality rate at 6 months. These findings support postoperative incentive spirometer use and show effectiveness of a simple intervention to improve incentive spirometer adherence.

Gamification

To demonstrate the real-time use of incentive spirometry data, iPad video game was created to be controlled by the add-on incentive spirometer device. In testing, the game showed no appreciable lag and continued connectivity during use, proving electronic spirometer data collected by the add-on device to be capable of real-time gamification applications. Gamification of medical interventions is an exciting concept for improving medical care adherence. Breathing games for the incentive spirometer is a familiar idea, with one group brainstorming a suite of games for asthmatics focusing on breathing metaphors as incentives for spirometer use [26], while others developed video games to incentivize breathing exercises and peak expiratory flow using digital flowmeters [18,27]. There exists an abandoned patent around the use of the incentive spirometer as a game controller [28] and an active patent around use of the flowmeter in video games [29], further supporting the popularity of the idea. The device in this study was created leveraging recent technologies and focusing design on clinical care use. Using the add-on device, as opposed to a digital flowmeter, maintains the current use of incentive spirometers in medical settings to allow native data capture. Potential reuse of the add-on device limits additional costs such as those incurred using digital flowmeters.

Limitations

Limitations to the add-on device design include contamination risks, costs, and technical and workflow implementation constraints. First, while completely enclosed, the add-on device is designed to be reused and carries the risk of infection—an especially important consideration in a respiratory pandemic such as COVID-19. To improve sterility, the device was enclosed in plastic, and sensors were placed behind the incentive spirometer, removing the need to expose the base to breaths from sensors placed below. Improvements can be made to close remaining gaps in the plastic encasing to further enclose the device and allow cleaning like an iPad or PlayStation controller, commonly used in the hospital settings. Second, routine incentive spirometer postoperative care has been estimated to carry a US $107.36 cost per patient above material cost of the spirometer device, which totals US $1.04 billion in total US annual costs [30]. This is a significant cost for a device with sparse evidence around use and poor patient compliance. While incentive spirometer devices are not reusable from patient to patient, the add-on device was designed to be reusable, lowering its effective cost. Overall, the add-on device carries a material price of approximately US $150 per unit (Multimedia Appendix 4), which could be reduced by bulk purchasing and further investigation into alternative individual components. While add-on device cost is an addition to the already significant price, identifying compliance and improving adherence will facilitate improved use and function of the incentive spirometer. Further studies of the incentive spirometer are required to investigate the prevention of postoperative breathing complications and their associated health care costs. These studies are dependent upon accurate compliance data and would benefit from the capabilities of the add-on device. Third, there exist integration and maintenance requirements of the add-on device, and it may be feasible for use only in hospital systems with existing technical support structures. The device was designed to minimize technical requirements, but more investigation is required. Future studies are required to trial the add-on device in clinical settings and test for improved adherence using strategies such as gamification compared to traditional incentive spirometry.

Conclusions

Incentive spirometers are routinely used in hospital settings, specifically in postoperative clinical care, but recommendations for proper routine use lack thorough investigation due to a general lack of data on device use. Creating a low-cost, effective, and reusable add-on device for the incentive spirometer allows native collection of previously inaccessible incentive spirometer compliance data. These data can facilitate research into incentive spirometer use to guide clinical care, incentivize adherence, and draw conclusions about the clinical effectiveness of the incentive spirometer.

Acknowledgments

The authors gratefully acknowledge the valuable contribution of Mikele Garrett and dedicate this paper in his memory. Mikele was an incredible person and teammate who passed away prior to the completion of this paper. He was critical in device development, overseeing and creating the add-on device designs. The authors also acknowledge the Adult Clinical Research Group at the University of Michigan for the valuable contribution to protocol development and final paper review; the University of Michigan Hacks with Friends competition for initial idea development; and the University of Michigan Software Development for Accessibility course EECS 495 and its students for video game ideas and creation, Amanda Bahamonde for her assistance with preparing this paper for submission, Ruth Cassidy for statistical support, and Kent Miller for access to the digital flowmeter used in device testing. Partial funding can be attributed to the Department of Anesthesiology, University of Michigan Medical School, the Society for Technology in Anesthesia 2021 Neurowave Research Grant, and the National Institute for General Medical Sciences of the National Institutes of Health (MLB; grant T32GM103730).

Data Availability

Deidentified data were collected from spirometer use by the study team. Data were collected using the incentive spirometer, add-on device, and digital flowmeter. Data sets generated and analyzed within this study are available from the corresponding author upon reasonable request.

Conflicts of Interest

This work has been declared through the University of Michigan Innovation Partnerships.

Multimedia Appendix 1

Add-on device design and schematics.

DOCX File , 294 KB

Multimedia Appendix 2

Alternative design schematic for add-on device.

DOCX File , 436 KB

Multimedia Appendix 3

Video displaying an investigator (TM) using the incentive spirometer with the add-on device to control the created video game.

MP4 File (MP4 Video), 7764 KB

Multimedia Appendix 4

Add-on device cost.

DOCX File , 13 KB

  1. Rock P, Rich PB. Postoperative pulmonary complications. Curr Opin Anaesthesiol. 2003;16(2):123-131. [CrossRef] [Medline]
  2. Fisher BW, Majumdar SR, McAlister FA. Predicting pulmonary complications after nonthoracic surgery: a systematic review of blinded studies. Am J Med. 2002;112(3):219-225. [CrossRef] [Medline]
  3. Gatti G, Cardu G, Lusa AM, Pugliese P. Predictors of postoperative complications in high-risk octogenarians undergoing cardiac operations. Ann Thorac Surg. 2002;74(3):671-677. [FREE Full text] [CrossRef] [Medline]
  4. Carter AR, Sostman HD, Curtis AM, Swett HA. Thoracic alterations after cardiac surgery. AJR Am J Roentgenol. 1983;140(3):475-481. [FREE Full text] [CrossRef] [Medline]
  5. Bartlett RH, Krop P, Hanson EL, Moore FD. Physiology of yawning and its application to postoperative care. Surg Forum. 1970;21:222-224. [Medline]
  6. Bartlett RH, Gazzaniga AB, Geraghty TR. Respiratory maneuvers to prevent postoperative pulmonary complications. A critical review. JAMA. 1973;224(7):1017-1021. [Medline]
  7. Franklin E, Anjum F. Incentive Spirometer and Inspiratory Muscle Training. Treasure Island, FL. StatPearls Publishing; 2022.
  8. Pullen RL. Teaching bedside incentive spirometry. Nursing. 2003;33(8):24. [CrossRef] [Medline]
  9. Martin TJ, Patel SA, Tran M, Eltorai AS, Daniels AH, Eltorai AEM. Patient factors associated with successful incentive spirometry. R I Med J (2013). 2018;101(9):14-18. [Medline]
  10. Strickland SL, Rubin BK, Drescher GS, Haas CF, O'Malley CA, Volsko TA, et al. American Association for Respiratory Care‚ Irving‚ Texas. AARC clinical practice guideline: effectiveness of nonpharmacologic airway clearance therapies in hospitalized patients. Respir Care. 2013;58(12):2187-2193. [FREE Full text] [CrossRef] [Medline]
  11. Freitas RFS, Soares GOS, Cardoso RC, Atallah ÁN. Incentive spirometry for preventing pulmonary complications after coronary artery bypass graft. Cochrane Database Syst Rev. 2012;2012(9):CD004466. [FREE Full text] [CrossRef] [Medline]
  12. Eltorai AEM, Baird GL, Eltorai AS, Healey TT, Agarwal S, Ventetuolo CE, et al. Effect of an incentive spirometer patient reminder after coronary artery bypass grafting: a randomized clinical trial. JAMA Surg. 2019;154(7):579-588. [FREE Full text] [CrossRef] [Medline]
  13. Amin R, Alaparthi GK, Samuel SR, Bairapareddy KC, Raghavan H, Vaishali K. Effects of three pulmonary ventilation regimes in patients undergoing coronary artery bypass graft surgery: a randomized clinical trial. Sci Rep. 2021;11(1):6730. [FREE Full text] [CrossRef] [Medline]
  14. Malik PRA, Fahim C, Vernon J, Thomas P, Schieman C, Finley CJ, et al. Incentive spirometry after lung resection: a randomized controlled trial. Ann Thorac Surg. 2018;106(2):340-345. [FREE Full text] [CrossRef] [Medline]
  15. Restrepo RD, Wettstein R, Wittnebel L, Tracy M. Incentive spirometry: 2011. Respir Care. 2011;56(10):1600-1604. [FREE Full text] [CrossRef] [Medline]
  16. Narayanan ALT, Hamid SRGS, Supriyanto E. Evidence regarding patient compliance with incentive spirometry interventions after cardiac, thoracic and abdominal surgeries: a systematic literature review. Can J Respir Ther. 2016;52(1):17-26. [FREE Full text] [Medline]
  17. Eltorai AEM, Baird GL, Eltorai AS, Pangborn J, Antoci V, Cullen HA, et al. Incentive spirometry adherence: a national survey of provider perspectives. Respir Care. 2018;63(5):532-537. [FREE Full text] [CrossRef] [Medline]
  18. Lange B, Flynn S, Rizzo A, Bolas M, Silverman M, Huerta A. Breath: a game to motivate the compliance of postoperative breathing exercises. Presented at: 2009 Virtual Rehabilitation International Conference; June 29, 2009-July 02, 2009, 2009; Haifa, Israel. [CrossRef]
  19. Ashton K. That ‘internet of things’ thing. Radio Freq Identif J. 2009;22(7):97-114. [FREE Full text]
  20. Yin Y, Zeng Y, Chen X, Fan Y. The internet of things in healthcare: an overview. J Ind Inf Integration. 2016;1:3-13. [CrossRef]
  21. Dimitrov DV. Medical internet of things and big data in healthcare. Healthc Inform Res. 2016;22(3):156-163. [FREE Full text] [CrossRef] [Medline]
  22. Fekr AR, Radecka K, Zilic Z. Design and evaluation of an intelligent remote tidal volume variability monitoring system in E-health applications. IEEE J Biomed Health Inform. 2015;19(5):1532-1548. [CrossRef] [Medline]
  23. Janidarmian M, Fekr AR, Radecka K, Zilic Z. A comprehensive analysis on wearable acceleration sensors in human activity recognition. Sensors (Basel). 2017;17(3):529. [FREE Full text] [CrossRef] [Medline]
  24. Patel S, Park H, Bonato P, Chan L, Rodgers M. A review of wearable sensors and systems with application in rehabilitation. J Neuroeng Rehabil. 2012;9:21. [FREE Full text] [CrossRef] [Medline]
  25. Park T, Hwang I, Lee U, Lee SI, Yoo C, Lee Y, et al. ExerLink: enabling pervasive social exergames with heterogeneous exercise devices. Presented at: MobiSys '12: Proceedings of the 10th International Conference on Mobile Systems, Applications, and Services; June 25-29, 2012, 2012;15-28; Low Wood Bay Lake District, UK. [CrossRef]
  26. van Delden R, Bos DPO, de With AJV, Vogel K, Klaassen R, Zwart N, et al. SpiroPlay, a suite of breathing games for spirometry by kids & experts. Presented at: CHI PLAY '20: Proceedings of the Annual Symposium on Computer-Human Interaction in Play; November 2-4, 2020, 2020;400-413; Virtual Event, Canada. [CrossRef]
  27. Chelabi K, Balli F, Bransi M, Gervais Y, Marthe C, Tse SM. Validation of a portable game controller to assess peak expiratory flow against conventional spirometry in children: cross-sectional study. JMIR Serious Games. 2021;9(1):e25052. [FREE Full text] [CrossRef] [Medline]
  28. Elefteriades J, Bango J, Dziekan M. Game-based incentive spirometer and a method of quantifying and recording performance. US Patent. URL: https://patentimages.storage.googleapis.com/b6/69/8c/ea18a2f81ebf10/US20130303930A1.pdf [accessed 2022-12-28]
  29. Luttrell RS. Respiratory therapy instrument offering game-based incentives, training, and telemetry collection. US Patent. 2019. URL: https://patentimages.storage.googleapis.com/60/ba/5c/51478d0167ed70/US10506950.pdf [accessed 2022-12-28]
  30. Eltorai AEM, Baird GL, Pangborn J, Eltorai AS, Antoci V, Paquette K, et al. Financial impact of incentive spirometry. Inquiry. 2018;55:46958018794993. [FREE Full text] [CrossRef] [Medline]


API: application programming interface


Edited by T Leung; submitted 20.02.23; peer-reviewed by P McCormick, J Mandel; comments to author 16.06.23; revised version received 07.07.23; accepted 27.07.23; published 07.09.23.

Copyright

©Michael L Burns, Anik Sinha, Alexander Hoffmann, Zewen Wu, Tomas Medina Inchauste, Aaron Retsky, David Chesney, Sachin Kheterpal, Nirav Shah. Originally published in JMIR Biomedical Engineering (http://biomsedeng.jmir.org), 07.09.2023.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Biomedical Engineering, is properly cited. The complete bibliographic information, a link to the original publication on https://biomedeng.jmir.org/, as well as this copyright and license information must be included.