Abstract
This pilot study found that an artificial intelligence–supported fingerprint-based patient identification system demonstrated preliminary technical feasibility and user acceptability among healthcare workers in a simulated clinical environment, warranting further real-world evaluation.
JMIR Biomed Eng 2026;11:e88657doi:10.2196/88657
Keywords
Introduction
An estimated 1 in 20 patients is exposed to preventable harm, with 12% of harm considered severe or fatal []. Patient identification errors are the root cause of many patient safety incidents in both inpatient and outpatient settings [,].
Common patient identification methods include person-specific information printed on wristbands or a person’s medical card. Additionally, algorithmic approaches, matching software, radio frequency identification devices, and biometrics may be used. All patient identification methods have limitations to operability []. A primary goal of the Joint Commission [] is to improve the accuracy of patient identification. Innovative approaches are needed to accomplish this goal.
Fingerprints are individual, nontransferable, and inherent to each person, thus ideal for patient identification. Heuristik utilizes advanced fingerprint identification technology () supported by artificial intelligence to enable identification in adverse situations (wet or dirty fingerprints). Upon admission, consenting patients can opt for fingerprint collection using Heuristik’s biometric software instead of standard identification procedures. Using secure connectivity, the encrypted fingerprints are then sent to a secure cloud-based server for remote processing, where they are pseudonymized and stored. To ensure privacy, Heuristik creates a unique identification method linked to the fingerprint. This method acts as a secure bridge to the patient’s medical record without accessing the patient’s medical record information.
This study investigates the technical feasibility (accuracy) and acceptability (user satisfaction) of the Heuristik system.

Methods
Study Design
A simulated multimethods cross-sectional study was conducted in March 2025 at Hospital da Luz’s simulation center in Lisbon, Portugal. The primary objective was to evaluate technical feasibility by matching fingerprints against a pseudonymized database across simulated scenarios. For privacy, Heuristik operated standalone without live electronic health records integration. Participants recruited via purposeful sampling included nurses, technicians, and other health care workers. Eligibility required being >18 years old, providing informed consent, and having usable fingerprints.
The workflow involved a brief training session, then a one-time enrollment phase prior to admission, followed by identification scans at preoperative, postoperative, and outpatient settings. Four Heuristik devices were deployed across scenarios. Before admissions, 12 simulated patients were enrolled using two fingerprints per hand. Researchers used structured forms to manually calculate accuracy by comparing point-of-care scans against registration data.
Usability data were collected via a revised version of the 10-item System Usability Scale tool (SUS) [], where scores can range from 0‐100 (68 considered as average []). Six items were added to the tool to assess usefulness, reliability, relevance, and error reduction on a 5-point scale, plus two open-ended items regarding likes and dislikes. Scale items underwent descriptive analysis in Excel; two researchers content-analyzed open-ended responses.
Ethical Considerations
The study was approved by the Luz Saúde Investigation Committee (February 20, 2025) and the Hospital da Luz Lisboa Ethics Committee (March 14, 2025; Ref. CES/11/2025/IE). The research adhered to the Helsinki Declaration, and all participants provided written informed consent prior to enrollment.
Results
Eighteen health care workers with a mean age of 36 (range 24‐50 y, 56% female, 100% Portuguese) participated in the study. The cohort was divided into two groups: 9 health care professionals acting as system testers (5 nurses, 2 patient-care technicians, and 2 client-service technicians) and 9 nonclinical participants ('others’) who acted as simulated patients. Within the professional group, 3 nurses assumed dual roles as both testers and patients, resulting in a total of 12 simulated patients and 6 professional participants who acted solely as system users.
Immediately prior to the clinical workflows, the 12 simulated patients were enrolled using two fingerprints per hand. Subsequent verification confirmed 100% identification accuracy for all participants using a scan of at least one enrolled fingerprint. Upon admission to the simulated settings, client service technicians independently scanned two fingerprints from two different simulated patients. Each patient underwent three scans using alternating hands, totaling six successful identifications with a 100% match rate. A team of six staff members (four nurses and two technicians) performed all subsequent identifications on four patients using bilateral scans. All the scans performed across three clinical settings (outpatient, preoperative, and postoperative) achieved 100% accuracy.
Regarding the acceptability data, seven staff members completed the survey. The SUS scores ranged from 70 to 97.5 (mean 87.14, SD 8.28). All (n=7; 100%) rated the 5-point scale items as “strongly agree” or “agree”. Open-ended responses revealed that participants liked the “reliability” (n=3), “ease of use or practical” (n=3), and “functionality and speed of client identification” (n=1) most about the product. Responses to “What did you like least about the product” included “nothing” (n=2) or were unable to be analyzed (n=5) as the responses did not refer to the product. These results suggest that Heuristik demonstrates high technical accuracy and user acceptability in a controlled simulated environment.
Discussion
While preliminary technical feasibility and high staff acceptability suggest the system addresses critical safety protocol gaps, these findings require cautious interpretation. Key limitations include small sample size, which precludes definitive claims regarding large-scale clinical efficacy, reliance on simulated settings lacking real-world clinical pressures, and the absence of operational metrics. Moreover, due to manual data collection limitations, granular biometric metrics—such as initial scan attempts, False Rejection Rate, and False Acceptance Rate—were not captured but remain a primary focus for future real-world studies. Furthermore, while staff adoption was high, patient acceptance remains a critical barrier often fueled by data privacy anxieties and opaque storage practices []. Future deployments must prioritize transparent communication to foster trust, as structured communication and risk assessments significantly improve staff acceptance and project success [,]. Finally, future real-world trials must validate safety improvements by assessing the system against damaged fingerprints, environmental variability, and infrastructure dependencies.
In conclusion, this pilot demonstrates preliminary technical feasibility and user acceptability. While these findings do not measure hard clinical outcomes like workflow efficiency or real-world error reduction, they establish foundational technical reliability, warranting further in vivo studies. Anticipated improvements in patient safety and clinical operations require empirical validation through planned real-world clinical studies.
Acknowledgments
Generative AI Disclosure
During the preparation of this work, the authors used Gemini (Google) strictly for language refinement and word-count reduction to meet formatting requirements. After using this tool, the authors rigorously reviewed and edited the content as needed.
Funding
This work was supported by the EIT Health Living Labs and Test Beds Programme.
Conflicts of Interest
GS was financially compensated as an independent consultant to assist in the preparation and submission of this manuscript. The author declares no financial ownership, equity interest, or other competing interests related to the subject matter of this research beyond the consulting compensation received. MD declares a conflict of interest as an employee of Heuristik Technologies, the company that developed the technology investigated in this study. AC declares a conflict of interest as the Founder, Chief Executive Officer (CEO), and a shareholder of Heuristik Technologies, the company that developed the technology investigated in this study
References
- Panagioti M, Khan K, Keers RN, et al. Prevalence, severity, and nature of preventable patient harm across medical care settings: systematic review and meta-analysis. BMJ. Jul 17, 2019;366:l4185. [CrossRef] [Medline]
- De Rezende HA, Melleiro MM, Shimoda GT. Interventions to reduce patient identification errors in the hospital setting: a systematic review protocol. JBI Database System Rev Implement Rep. Jan 2019;17(1):37-42. [CrossRef] [Medline]
- Kulju S, Morrish W, King L, Bender J, Gunnar W. Patient misidentification events in the Veterans Health Administration: a comprehensive review in the context of high-reliability health care. J Patient Saf. Jan 1, 2022;18(1):e290-e296. [CrossRef] [Medline]
- Riplinger L, Piera-Jiménez J, Dooling JP. Patient identification techniques - approaches, implications, and findings. Yearb Med Inform. Aug 2020;29(1):81-86. [CrossRef] [Medline]
- National Patient Safety Goals. The Joint Commission. 2025. URL: https://www.jointcommission.org/en-us/standards/national-patient-safety-goals [Accessed 2026-06-29]
- Brooke J. SUS-a quick and dirty usability scale. In: Usability Evaluation in Industry. Vol 189. Taylor & Francis; 1996:4-7. URL: https://www.taylorfrancis.com/chapters/edit/10.1201/9781498710411-35/sus-quick-dirty-usability-scale-john-brooke [Accessed 2026-06-29]
- Lewis JR. The System Usability Scale: past, present, and future. Int J Hum -Comput Interact. Jul 3, 2018;34(7):577-590. [CrossRef]
- Biometric technology in healthcare: balancing security benefits with implementation challenges. World J Adv Res Rev. 2025;26(1):1864-1870. [CrossRef]
- Sony M, Antony J, Tortorella GL. Critical success factors for successful implementation of Healthcare 4.0: a literature review and future research agenda. Int J Environ Res Public Health. Mar 6, 2023;20(5):4669. [CrossRef] [Medline]
- Hilda H. What is biometric authentication and how to implement it? HyperVerge; 2024. URL: https://hyperverge.co/blog/biometric-authentication/ [Accessed 2026-06-29]
Abbreviations
| SUS: System Usability Scale |
Edited by Javad Sarvestan; submitted 30.Nov.2025; peer-reviewed by Erika Ramos, Stephen Kuliju; final revised version received 08.Jun.2026; accepted 10.Jun.2026; published 09.Jul.2026.
Copyright© Ginger Schroers, Marta del Rio Oliva, Jenny O'Rourke, Antxon Caballero, Luis Barata Gomes, Raquel Araújo. Originally published in JMIR Biomedical Engineering (http://biomsedeng.jmir.org), 9.Jul.2026.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Biomedical Engineering, is properly cited. The complete bibliographic information, a link to the original publication on https://biomedeng.jmir.org/, as well as this copyright and license information must be included.